Carolyn:
Although various "citrate based" AR formulations work fairly well. We have
found that Novocastra's 6F11 works much better when using a 0.1 M Tris-HCl
pH 9. See Brancroft and Stevens for proper ratios of Tris Base and
Tris-HCl.
Hope this helps,
Carlos Genty
Breast Center Pathology
Baylor College of Medicine
Houston, Texas
cgenty@bcm.tmc.edu