Re: [Histonet] Ventana Benchmark XTs

From:"Piero Nelva"

Hi Tracey

We had exactly that problem with inconsistent slide temps and failure to 
heat on random pads.

Our local service technician said there was a manufacturing problem with all 
plates made during a certain time and that they all needed to be replaced 
worldwide.  We had our pad tray replaced about a month ago and not a problem 
since.  Perhaps you can check if your replacement tray was from the "new" 


Piero Nelva

----- Original Message ----- 
From: "Tracey Lenek" 
Cc: "Martin Trotter" ; "Joanna Bartczak-McKay" 

Sent: Thursday, November 27, 2008 3:59 AM
Subject: [Histonet] Ventana Benchmark XTs


We have had on-going thermal slide pad issues with our XTs since they were 
in February. The slide tray assemblies have been replaced not once but twice 
and we are still having inconsistent
results with the temp verifier slides.  Has anyone experienced the same 
issues with this

Tracey Lenek
Tech III - Anatomic Pathology
Calgary Laboratory Services

This message and any attached documents are only for the use of the intended 
recipient(s), are confidential and may contain privileged information. Any 
unauthorized review, use, retransmission, or other disclosure is strictly 
prohibited. If you have received this message in error, please notify the 
sender immediately, and then delete the original message. Thank you.