[Histonet] Stainng Problems-Ventana -- Benchmark

From:"Botsford, Daniel"

Re Robert Bates florida

Who manufactures your positive charged slide?

We had ER cases where the positive control on the top of the slide stained
and the a known  positive patient case on the bottom half did not. Internal
control did not stain. Traced the problem to slide manufactor. 
Manufactured failed to have an even distribution of charge on the entire
slide. Use Erie Slides

Sincerely

Daniel Botsford
Windsor Regional Hospital
1995 Lens Avenue
Windsor, Ontario
N8W 1L9
519-254-5577 ext 52373
519-254-6861 fax