RE: [Histonet] DAKO ER and PR antibodies

From:Rena Fail

Try 1:10 for ER and 1:150 for PR with HIER pH 6.0 20 minutes steam, 10
min cool down
Rena Fail

-----Original Message-----
[] On Behalf Of THERESA
Sent: Thursday, October 02, 2003 11:00 AM
Subject: [Histonet] DAKO ER and PR antibodies

Hi out there in Histonet land:

I am having a problem working up dilutions for DAKO's ER (M7047) and PR 
(M3569) antibodies.
I am using the Envision+ Detection system

Suggestions from DAKO were:

ER  1:100 to 1:150 with Envision+ and a 20 minute target retrieval PR
1:200 to 1:400 with Envision+ and a 20 minute target retrieval.

I've gone less and more and still cannot get any where near the staining
these cases that went out to a IHC Laboratory than I am using as a 

Any suggestions would certainly be appreciated.