We have been routinely using the ckit from Dako, and have not seen this type
of staining. We have compared this antibody to a couple of other clones, and
it has continuously come out on top. I am wondering if perhaps either
tissues fixation/processing is affecting the staining. Or perhaps it is a
heat pretreatment or titer issue. Just my thoughts.
Patti Loykasek
Phenopath Laboratories
Seattle, WA>
One of our pathologists has shared some information he was given by a
doctor
> at AFIP. I have never seen this before, but was wondering if anyone else
> has.
>
> They claim that the c-Kit antibody from Dako(A4502)gives "widespread
> positivty of fibroblasts and myofibroblasts, suggesting non-specific cross
> reactivity with this particular antibody". However, they did not experience
> this phenomenon with the c-kit from Santa Cruz Biotechnology (sc-168). Dako
> are not aware of any issues with their antibody.
>
> Both of these are polyclonals.
>
> Appreciate your input
> Thanks
> Ronnie
>