We have a Tissue Tek VIP that is about 13 years old and a Leica TP 1050 that
is about 3 years old.  We have noticed on both processors the smell of
formalin when the processing run was complete.  I am going to check the
vapor trap bottles.  Thanks!

Laurie

-----Original Message-----
From: Medilex123@aol.com [mailto:Medilex123@aol.com]
Sent: Wednesday, March 22, 2000 8:17 AM
To: LColbert@phsca.org; Histonet@pathology.swmed.edu
Subject: Re: FW: H & E Staining problems


The Hypercenter 2 and the early vesions of the Hypercenter XP. We havn't had

a serious problem since Shandon upgraded the valves on our 3 XPs (at their 
own, considerable expense) The Leica doesn't appear to be prone to valve 
failure (a colleague uses one) but the same colleague had a problem when its

vapour trap bottle became full and backed up formalin and other fluids into 
the last xylene and antemedia bottle. Routine maintenance was not being 
performed. 
I think the vapour trap exists on all enclosed processors, its worth
checking 
to see if its full.
Which one do you use?


Glyn