Ventana 1530 tissue processor

From:"Featherstone, Annette"

Re: HE stain without Blue reagent
Hi
We are in dire need of a manual for the Ventana 1530 tissue processor. Does anyone out there have it and could I coerce you into sending me a copy? We would be forever indebted!!
Annette Featherstone HT/MLT, Kaleida Health, Buffalo NY
 

-----Original Message-----
From: Joey Shepherd [mailto:j.shepherd@sheffield.ac.uk]
Sent: Thursday, June 05, 2003 11:43
To: histonet@pathology.swmed.edu
Subject: Re: HE stain without Blue reagent

Hi Yan
You should be able to blue your sections just by running under tap water, but you can try making your own blueing solution - I use Scott's solution (20 g/l of sodium bicarbonate and 3.5 g/l of magnesium sulphate 7-hydrate). After haeamtoxylin, rinse slides in tap water, then immerse in Scott's for 30s-1min, then rinse in tap water before eosin.
Joey.
-- 