RE: DAKO autostainer

From:Rena Fail <RFail@charleston.net>

<html> <font size=3>THANK YOU PATSY!<br> <br>           Just the info I needed for initial dilutions. trials. <br> Rena<br> <br> <br> At 09:29 PM 6/8/01 -0600, you wrote:<br> <blockquote type=cite class=cite cite>envision is much more inhanced than a lsab detection system, i would dilute<br> the antibodies used with the lsab system, at least 10 times, so if you used<br> an antibody at 1:100 with lsab, use it at 1:1000 with envision and for maybe<br> less time.<br> patsy ruegg <br> <br> -----Original Message-----<br> From: Rena Fail<br> To: histonet@pathology.swmed.edu<br> Sent: 6/7/01 4:39 PM<br> Subject: DAKO autostainer<br> <br> We just purchased the DAKO autostainer. I would really appreciate any <br> advise on titrations of antibodies with the envision plus kit and alk<br> phos.<br> <br> Thanks<br> Rena Fail, AS,HT(ASCP)<br> SS/IHC<br> Medical University of SC<br> Charleston, SC</font></blockquote><br> </html>