At 09:29 PM 6/8/01 -0600, you wrote:<br>
<blockquote type=cite class=cite cite>envision is much more inhanced than
a lsab detection system, i would dilute<br>
the antibodies used with the lsab system, at least 10 times, so if you
used<br>
an antibody at 1:100 with lsab, use it at 1:1000 with envision and for
maybe<br>
less time.<br>
patsy ruegg <br>
<br>
-----Original Message-----<br>
From: Rena Fail<br>
To: histonet@pathology.swmed.edu<br>
Sent: 6/7/01 4:39 PM<br>
Subject: DAKO autostainer<br>
<br>
We just purchased the DAKO autostainer. I would really appreciate any
<br>
advise on titrations of antibodies with the envision plus kit and
alk<br>
phos.<br>
<br>
Thanks<br>
Rena Fail, AS,HT(ASCP)<br>
SS/IHC<br>
Medical University of SC<br>
Charleston, SC</font></blockquote><br>