RE: Benchmark error code

From:"Sebree Linda A."


We've experienced this error message on our Discovery and after replacing some valves found that its actually caused by sediment in the EZ Prep itself which clogs up the sieve on the intake tube.  We cleaned this screen and filtered the EZ Prep and so far have had no further problem.

Linda A. Sebree, HT(ASCP)
University of Wisconsin Hospital & Clinics
IHC/ISH Laboratory
600 Highland Ave.
Madison, WI 53792
FAX: (608)262-7174

-----Original Message-----
From: Lynn McKee []
Sent: Wednesday, July 24, 2002 11:01 AM
Subject: Benchmark error code

Dear Histonetters,

    We're curious to see if anyone with a Benchmark experienced the error
code "EZ Prep resevoir fails to charge" ? 

Lynn McKee
University of Cincinnati

<< Previous Message | Next Message >>