Dear Inga,

On our cryostats I would have the specimen temperature set @ -12C and the microtome chamber temperature set @ -20C as this will stop any adherence to the knife and anti-roll plate. You are much too cold at present on the specimen temperature. 
I see you are a neuro lab these settings would be fine for brain too.

Best Regards

Alan Bright