Re: [Histonet] New Microwave Cap Regulation

From:Phil McArdle

Here's a "vendor" reply:

First of all, CAP's annual leakage measurement requirement was deleted 
12.12.06.

Second, should you still desire to check for leakage (always a good 
idea), a local appliance repair shop should be able to do it, but they 
may not wish to from a liability standpoint.

There are also many sources for leakage measurement devices you can use 
yourself. The cheapest are prone to false positives; avoid "passive" 
type devices (don't require a battery) and "idiot-light" type. Expensive 
devices that "sum" and have built-in measurement spacers are available, 
but we feel are "overkill" for this application. Apologies for the 
following "shameless commerce," but EBS offers a "good enough" active 
(battery powered) handheld unit that provides a digital readout (model 
H2501) for $53, with  instructions included that are tailored to a 
pathology lab context.

Best regards,

Phil McArdle

-- 
Microwave Product Manager

Energy Beam Sciences, Inc.
29-B Kripes Rd.
East Granby, CT 06026

Tel:  800.992.9037 x 341
Fax: 860.653.0422

pmcardle@ebsciences.com
www.ebsciences.com

You must be the change you want to see in the world.
- Mahatma Gandhi

NOTE: This message, together with any attachments, is intended only for the use of the individual or entity to which it is addressed and may contain information that is legally privileged, confidential and exempt from disclosure. If you are not the intended recipient, however, there's not a lot I can do about it, and it was probably my mistake anyway; please do the right thing and make this e-mail go away. Thank you.


Stacey Burton wrote:
> Thank you Hector.
> This brings me to a new question - How do I test for leakage is less than 5mW/cm2 at a distance of 5 cm from the surface?  Or what type of service to I call to help me with that?
> -Stacey
>
>
> ----- Original Message ----
> From: Hector Hernandez 
> To: Rene J Buesa ; "Luck, Greg D." ; Stacey Burton ; histonet@lists.utsouthwestern.edu
> Sent: Thursday, February 22, 2007 4:19:54 PM
> Subject: RE: [Histonet] New Microwave Cap Regulation
>
>
> ANP.29430   Phase 1
> This question doesn't state laboratory or home grade microwaves.  Microwave
> should be placed in an appropriate ventilated hood to contain airborne
> chemical contaminates and potential infectious agents. Microwaves used
> outside a fume hood should have an integral fume extractor that is certified
> by the manufacturer for use in a clinical laboratory.  This doesn't say
> microwave has to be laboratory grade, only the fume extractor. This
> checklist question doesn't apply if only non-hazardous reagents are used in
> the device (microwave) (e.g. water, certain biological stains)  
>
> In my opinion you can use any microwave (laboratory or house microwave) as
> long as you can document:
> 1. Leakage is less than 5mW/cm2 at a distance of 5 cm from the surface.
> 2. Periodically monitor for temperature reproducibility.
> 3. All containers used in the microwave are made from microwave-transparent
> material/ microwave proof.
> Don't make it any harder than it is.
>
> Hope this clears a few things,
>
> Hector
>
>
> -----Original Message-----
> From: histonet-bounces@lists.utsouthwestern.edu
> [mailto:histonet-bounces@lists.utsouthwestern.edu] On Behalf Of Rene J Buesa
> Sent: Thursday, February 22, 2007 2:19 PM
> To: Luck, Greg D.; Stacey Burton; histonet@lists.utsouthwestern.edu
> Subject: RE: [Histonet] New Microwave Cap Regulation
>
> Greg:
>   The new CAP regulation stems from the fact that many labs used the "house
> microwaves ovens"(HMWO) also to fix and process tissues. That is where CAP,
> manufacturers and a MW task force intervened with the valid warning that
> fixing and processing, and even staining when the solutions are alcoholic,
> are unsafe procedures to be done in a regular
>   HMWO,and that those procedures really require an instrument with a venting
> capability that will take the noxious vapors out of the lab.
>   To just heat water or buffers for HIER I did that for many years and no
> inspector ever said anything. Perhaps they have become more demanding
> recently on this regulation, but for me there is nothing noxious to vent
> when only water, PBS or buffers are heated. In your house you also heat
> water and there is nothing noxious about it (and we are talking about the
> "sacred" home environment).
>   Just my opinion though!
>   René J.
>
> "Luck, Greg D."  wrote:
>   Hello all,
> Would venting or use of a "Laboratory Grade" microwave be required if
> all that it is used for is to heat water or PBS or Tris buffers?
> Thanks, Greg
> Greg Luck, B.S., HT(ASCP)
> Anatomic Pathology Supervisor
> Deaconess Med Cntr
> 800 W. 5th Ave
> Spokane, WA 99204
> Offc 509.473.7077
> Fax 509.473.7133
> luckg@empirehealth.org 
> www.deaconessmc.org
>
>
>
> -----Original Message-----
> From: histonet-bounces@lists.utsouthwestern.edu
> [mailto:histonet-bounces@lists.utsouthwestern.edu] On Behalf Of Stacey
> Burton
> Sent: Thursday, February 22, 2007 8:22 AM
> To: histonet@lists.utsouthwestern.edu
> Subject: [Histonet] New Microwave Cap Regulation
>
> I am curious to know how labs have responding to the CAP Question
> ANT.29430 which asks "Are microwave devices properly vented?".
> Have the labs continued to utilized their regular "kitchen model"
> microwaves and applied duct work to them to be vented out of the
> building?
> Please comment.
>
> Thank you,
> Stacey Burton, H.T. ASCP
> Precision Pathology Services
> San Antonio Texas
>
>
>
> ________________________________________________________________________
> ____________
> Food fight? Enjoy some healthy debate
> in the Yahoo! Answers Food & Drink Q&A.
> http://answers.yahoo.com/dir/?link=list&sid=396545367
> _______________________________________________
> Histonet mailing list
> Histonet@lists.utsouthwestern.edu
> http://lists.utsouthwestern.edu/mailman/listinfo/histonet
>
> _______________________________________________
> Histonet mailing list
> Histonet@lists.utsouthwestern.edu
> http://lists.utsouthwestern.edu/mailman/listinfo/histonet
>
>
>
>
>
> ---------------------------------
> Now that's room service! Choose from over 150,000 hotels 
> in 45,000 destinations on Yahoo! Travel to find your fit.
> _______________________________________________
> Histonet mailing list
> Histonet@lists.utsouthwestern.edu
> http://lists.utsouthwestern.edu/mailman/listinfo/histonet
>
>
>  
> ____________________________________________________________________________________
> Don't get soaked.  Take a quick peak at the forecast
> with the Yahoo! Search weather shortcut.
> http://tools.search.yahoo.com/shortcuts/#loc_weather
> _______________________________________________
> Histonet mailing list
> Histonet@lists.utsouthwestern.edu
> http://lists.utsouthwestern.edu/mailman/listinfo/histonet
>   



_______________________________________________
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet


<< Previous Message | Next Message >>