On pump readout in back of machine, we have this inf, our processor was
purchased in 1984 and still going strong!!
Red numbers = vacuum goes to 50 - 60 cm Hg.
Black numbers = pressure, should be 0.25 - 0.3, below 0.2 relief valve may
need adjustment.
For delicate tissues, we turn the pressure/vacuum off, or program it as
OFF/CLEAR for ALL STATION
Sakura has a way to do a pressure/vacuum check by keying in certain codes,
a phone call to their tech services would do this.