RE: [Histonet] microwaves in labs

From:"Bartlett, Jeanine (CDC/CCID/NCZVED)"



Since I started this whole thing let me reiterate my needs:  I was
simply asking if it would be a safety concern to heat a plastic coplin
jar of methenamine silver for a GMS for 24 seconds once  a week.  We
have a laboratory grade microwave but were looking to conserve space and
I just wondered if anyone had first hand experience with this very
narrow and specific limited use.

Thanks to everyone for their input. 


Jeanine Bartlett
Infectious Disease Pathology Branch
(404) 639-3590 
jeanine.bartlett@cdc.hhs.gov


-----Original Message-----
From: histonet-bounces@lists.utsouthwestern.edu
[mailto:histonet-bounces@lists.utsouthwestern.edu] On Behalf Of Phil
McArdle
Sent: Wednesday, August 22, 2007 1:31 PM
To: Histonet (E-mail)
Subject: Re: [Histonet] microwaves in labs


   And  now  a  laboratory  microwave vendor weighs in, so you can
decide
   whether  or not you want to read this, or what level of credibility
we
   have (I'm getting out my Nomex flame-retardant suit):
   First of all, you're not stuck paying $35k for a lab microwave, even
a
   model for processing.
   Second, please feel free to download our Microwave Companion, a
purely
   "advancing-the-art" publication that does not attempt to sell
anything
   at all: [1]http://www.ebsciences.com/pdf/EBS_MW_COMPANION.pdf
   Third, lab managers should be at least as concerned about OSHA as
CAP;
   see  OSHA 29 CFR 1910.303(b)(2) which simply states that no
food-grade
   microwave should be used for anything other than its intended
purpose.
   Now, here's essentially a repeat of an earlier Histonet posting:
   Energy Beam Sciences, a pioneer in the use of microwaves in
histology,
   has  maintained  for over a decade that consumer-grade microwaves
have
   no  place  in  a  laboratory  setting. Admittedly, over the years
some
   users  have  managed  to  achieve  varying  levels  of success in
some
   "non-tissue-processing" applications like staining. However, I'd
argue
   that  a  high number of failures due to this same approach have
served
   to  regress the art via anecdotal microwave "horror stories." Now
that
   microwaves  (both  consumer-grade and true laboratory
instrumentation)
   are  in  routine  and  widespread  use in the histology laboratory,
it
   should be unsurprising that CAP is taking this issue seriously.
   CAP  is  not  alone;  microwaves  in  histology  have  been
receiving
   increased  interest  from standards institutes such as CLSI (see
their
   publication GP-28A), and significantly, governmental agencies.
Besides
   Europe's   strict   IVDD   regulations,   in  the  USA  you  have
the
   aforementioned OSHA.
   It may be true that an external hood, fitted to a "kitchen"
microwave,
   will  remove  vapors  from the lab. But consumer grade microwaves
have
   other  limitations:  they are often prone to hot and cold spots
within
   their   chambers,  usually  aren't  designed  or  built  to  take
the
   day-in-and-day-out  demands  of  the  histology laboratory, don't
have
   calibrated  output,  and  don't  provide  the  means to measure
output
   power, for example. Due to their long magnetron cycle times, there
may
   be  no  difference  at  all  between  15  seconds at 100% power and
15
   seconds  at  20%  power.  These are not deficiencies, given the
proper
   context:  heating  leftovers  or  reheating  cold coffee at home...
an
   entirely different context from someone's biopsy, however.
   There's  also  the  question of liability exposure. If your
laboratory
   were  party to any kind of lawsuit, whether as plaintiff or
defendant,
   how  comfortable  would  you  be  if  it  came out that an
inexpensive
   consumer appliance were being used, rather than available,
appropriate
   laboratory  instrumentation,  even  if it had absolutely no bearing
on
   the issue at hand?
   Now how does that $69 Wall-to-Wall-Mart microwave stack up?
   Late  in  2005,  after  a series of conversations and e-mails with
CAP
   representatives  (and CLSI, whose publications CAP references),
Energy
   Beam  Sciences  developed  some recommendations to help with these
new
   checklist  items. An Acrobat file of said recommendations may be
found
   on our website at
   [2]http://www.ebsciences.com/pdf/EBS_CAP_RECOMMEND.pdf
   (Note  that  the  requirement  for annual leakage measurement has
been
   dropped, although it's still a good idea.)
   Also,  feel  free to browse our "library" section for articles,
papers
   and protocols.
   Although  I  do  all  I  can to sell our laboratory microwaves, I'm
at
   least  as interested in "advancing the art." To me, that translates
to
   getting   consumer   microwaves  out  of  labs,  and  true
laboratory
   microwaves in, even if they're not ours.
   Best regards,
   Phil McArdle
--
Phil McArdle
Microwave Product Manager

Energy Beam Sciences, Inc.
29-B Kripes Rd.
East Granby, CT 06026

Tel:  800.992.9037 x 341
Mobile: 860.597.6796
Fax: 860.653.0422

[3]pmcardle@ebsciences.com
[4]www.ebsciences.com

I skate to where the puck is going to be, not to where it's been.
- Wayne Gretsky

You must be the change you want to see in the world.
- Mahatma Gandhi

NOTE: This message, together with any attachments, is intended only for
the use  of the individual or entity to which it is addressed and may
contain informati on that is legally privileged, confidential and exempt
from disclosure. If you are not the intended recipient, however, there's
not a lot I can do about it, a nd it was probably my mistake anyway. So
please do the right thing and make thi s e-mail go away. Thank you.

   Parker, Helayne wrote:

     I  was  wondering if anyone new what CAP require of us if we have a
     common
     household microwave that we use for special stains.
     Helayne Parker,HT (ASCP)
     Histology Section Head
     Skaggs Community Health Center
     Branson, Missouri
     _______________________________________________
     Histonet mailing list
     [5]Histonet@lists.utsouthwestern.edu
     [6]http://lists.utsouthwestern.edu/mailman/listinfo/histonet
     _______________________________________________

   Histonet mailing list
   [7]Histonet@lists.utsouthwestern.edu
   [8]http://lists.utsouthwestern.edu/mailman/listinfo/histonet

   Bartlett, Jeanine (CDC/CCID/NCZVED) wrote:

Hi all:

I'd like to know how many of you out there use non-laboratory grade
microwaves for specials stains.......specifically for heating
methenamine-silver for a GMS. For those that do, are there any safety
issues of which you are aware?

Thanks for your help!

Jeanine Bartlett, BS, HT(ASCP)QIHC
Centers for Disease Control and Prevention Infectious Diseases Pathology
Branch 1600 Clifton Road, MS/G-32
18/SB-114
Atlanta, GA  30333
(404) 639-3590
[9]jeanine.bartlett@cdc.hhs.gov

_______________________________________________
Histonet mailing list
[10]Histonet@lists.utsouthwestern.edu
[11]http://lists.utsouthwestern.edu/mailman/listinfo/histonet

References

   1. http://www.ebsciences.com/pdf/EBS_MW_COMPANION.pdf
   2. http://www.ebsciences.com/pdf/EBS_CAP_RECOMMEND.pdf
   3. mailto:pmcardle@ebsciences.com
   4. http://www.ebsciences.com/
   5. mailto:Histonet@lists.utsouthwestern.edu
   6. http://lists.utsouthwestern.edu/mailman/listinfo/histonet
   7. mailto:Histonet@lists.utsouthwestern.edu
   8. http://lists.utsouthwestern.edu/mailman/listinfo/histonet
   9. mailto:jeanine.bartlett@cdc.hhs.gov
  10. mailto:Histonet@lists.utsouthwestern.edu
  11. http://lists.utsouthwestern.edu/mailman/listinfo/histonet
_______________________________________________
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet



_______________________________________________
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet




<< Previous Message | Next Message >>