Donna Carr wrote:
>
> Dear Fellow Histonetters,
> We are currently experiencing problems with our PAS/DPAS stain, We
> have replaced the Schiff reagent. And have tried different controls. I
> would greatly appreciate any advice on what tissues you use as controls
> and methods of staining. This is a recent problem that leads us to
> believe it is either a contaminated reagent or bad control. Any input
> is welcome. Thanks Donna
Hi Donna,
We use autopsy liver tissue as a control. Just verify the presence of
glycogen with a known control before use, not all liver tissues have
detectable amount of glycogen present.
Katri